This case report describes multimodal intraoperative neurophysiological monitoring during endoscopic lumbar discectomy in a 76-year-old woman with lumbar spinal canal stenosis and lumbar disc herniation, focusing on SSEP, MEP, EMG, and real-time nerve-root protection.
A 76-year-old woman was scheduled for endoscopic lumbar discectomy for lumbar spinal canal stenosis and lumbar disc herniation. Before surgery, she had pain and weakness in both lower extremities but remained able to walk.
2. Surgical Objective
The surgical objectives were to remove the herniated nucleus pulposus precisely under endoscopic visualization, release the compressed nerve root, achieve minimally invasive decompression, relieve the patient’s pain and weakness, and improve quality of life.
II. Intraoperative Neurophysiological Monitoring Methods and Procedures
The neurophysiological monitoring modalities commonly used during endoscopic lumbar discectomy include somatosensory evoked potentials (SEP), motor evoked potentials (MEP), and electromyography (EMG).
SEP was used to monitor the integrity of the dorsal columns and sensory conduction pathways.
MEP was used to monitor the anterior spinal cord, corticospinal tracts, and motor nerve-root function.
EMG was used primarily to detect direct stimulation of the nerve root in real time, provide an immediate intraoperative alert, and help avoid mechanical nerve injury.
The monitoring parameters, objectives, alert criteria, and intraoperative findings for each modality are described below.
1. Somatosensory Evoked Potentials (Lower-Extremity SSEP / LSEP)
1) Electrode Montage
Recording electrodes: The recording electrode was placed at the scalp Cz position, with Fz used as the reference electrode.
Stimulation electrodes: The bilateral posterior tibial nerves were stimulated at the ankles.
2) Recording and Stimulation Parameters
Recording parameters
| Parameter | Setting |
|---|---|
| Low-frequency filter | 30 Hz |
| High-frequency filter | 500 Hz |
| Sweep speed | 10 ms |
| Sensitivity | 2 microvolts |
Stimulation parameters
| Parameter | Setting |
|---|---|
| Stimulation frequency | 2.33 Hz |
| Pulse type | Monophasic pulse |
| Pulse width | 200 microseconds |
| Stimulation intensity | 25 mA, constant current |
3) Monitoring Points
Before surgery, the patient should be assessed for lower-extremity numbness, and any available preoperative EMG report should be reviewed. If available, the preoperative EMG examination may serve as a reference for intraoperative monitoring.
SEP should be monitored continuously throughout the operation, with attention to changes in waveform amplitude and latency. An amplitude reduction of at least 50% or a latency prolongation of at least 10% was described as indicating possible impairment of nerve-root conduction and requiring urgent decompression. Continuous monitoring also provides real-time warning of possible motor-pathway injury and is intended to reduce the risk of postoperative lower-extremity weakness, paralysis, or motor dysfunction.
The source case documented the following LSEP changes:
Preoperative baseline: A stable baseline waveform was obtained.
During endoscopic exploration: The right-sided somatosensory response decreased.
During nucleus-pulposus removal: The right-sided somatosensory response decreased.
During closure: The right-sided response increased and showed a trend toward recovery.

Stable preoperative lower-extremity SSEP waveform used to establish the baseline.

Right-sided lower-extremity SSEP decreased during endoscopic exploration.

Right-sided lower-extremity SSEP decreased during removal of the herniated nucleus pulposus.

The right-sided lower-extremity SSEP increased during closure, showing a trend toward recovery.
2. Motor Evoked Potentials (MEP)
1) Electrode Montage
Recording electrodes: Needle electrodes were inserted into muscles innervated by the L4-S1 nerve roots, including the quadriceps, tibialis anterior, gastrocnemius, and abductor hallucis muscles.
Stimulation electrodes: Transcranial electrical stimulation was applied over the motor cortex at the C1 and C2 positions.
2) Recording and Stimulation Parameters
Recording parameters
| Parameter | Setting |
|---|---|
| Low-frequency filter | 30 Hz |
| High-frequency filter | 3,000 Hz |
| Notch filter | Off |
| Sweep speed | 10 ms |
| Sensitivity | 50 microvolts |
Stimulation parameters
| Parameter | Setting |
|---|---|
| Pulse type | Monophasic pulse |
| Number of pulses per train | 5 |
| Pulse width | 75 microseconds |
| Train frequency | 1,000 Hz |
| Stimulation intensity | 300 V |
3) Monitoring Points
Before surgery, the patient should be assessed for lower-extremity weakness and muscle-strength grade.
Because MEP stimulation may cause patient movement, the surgeon must be instructed to stop the surgical maneuver before each intraoperative stimulation. The source case states that an SEP amplitude decrease of at least 30% or a latency prolongation of at least 10% requires intensified MEP monitoring. SEP findings should be interpreted together with the MEP waveform to assess whether neurological injury may be present.
MEP may also be monitored synchronously when intraoperative radiography is required. Disappearance or an abrupt decrease in MEP amplitude suggests possible acute injury to the motor pathway, and the hazardous maneuver should be terminated immediately.
The purpose of MEP monitoring is to provide real-time warning of motor-pathway injury and help prevent postoperative lower-extremity weakness, paralysis, and motor dysfunction.
The source case documented the following MEP stages:
Preoperative baseline: The baseline was recorded while neuromuscular blockade had not yet been completely metabolized.
During intraoperative radiography during endoscopic exploration: MEP was monitored after neuromuscular blockade had been metabolized.
During postoperative closure: The MEP waveform showed no substantial change compared with the waveform recorded during endoscopic exploration.

Preoperative MEP baseline recorded before complete metabolism of the neuromuscular blocking agent.

MEP monitoring during intraoperative radiography performed during endoscopic exploration, after neuromuscular blockade had been metabolized.

During closure, the MEP waveform showed no substantial change compared with the waveform recorded during endoscopic exploration.
3. Electromyography (EMG)
1) Electrode Montage
Recording electrodes: Recording electrodes were generally placed over the quadriceps, tibialis anterior, gastrocnemius, and abductor hallucis muscles.
2) Recording Parameters
| Parameter | Setting |
|---|---|
| Low-frequency filter | 30 Hz |
| High-frequency filter | 3,000 Hz |
| Sweep speed | 100 ms |
| Sensitivity | 50 microvolts |
3) Monitoring Objective
EMG was primarily used to monitor direct stimulation of the nerve root in real time, provide an immediate intraoperative alert, and help avoid mechanical nerve injury.
Burst or sustained EMG activity suggests that the nerve root may be undergoing mechanical stimulation, including contact, traction, or compression. When such activity appears, the surgical maneuver should be paused immediately and the position of the instrument adjusted.
The source case documented the following EMG findings:
Preoperative EMG: Baseline EMG signal.
During exploration: Burst EMG activity occurred.
During nucleus-pulposus removal: Burst EMG activity occurred.

Preoperative EMG signal used as the reference for intraoperative comparison.

Burst EMG activity during endoscopic exploration.

Burst EMG activity during removal of the herniated nucleus pulposus.
During puncture and insertion of the endoscope, the EMG signal showed resting electrical activity, and the SEP waveform remained stable.
Because the interval from induction to the beginning of this type of procedure was short and neuromuscular blockade had not yet been completely metabolized, the initial MEP amplitude was relatively low. A baseline was established to enable comparison during the remainder of the operation.

Under endoscopic visualization, the surgical team identified the superior articular process, ligamentum flavum, nerve root, dural sac, and herniated nucleus pulposus. Disc-space decompression was performed, and the herniated nucleus pulposus protruding into the spinal canal was extracted.
During traction on the herniated nucleus pulposus, the right-sided SEP amplitude decreased by approximately 10%-20%. Because neuromuscular blockade had continued to metabolize, the MEP amplitude was higher than the preoperative value. Marked burst EMG activity was observed in the left tibialis anterior muscle during traction on the herniated nucleus pulposus.


During nerve-root decompression and removal of the herniated nucleus pulposus, the left-sided SEP showed no substantial change, while the right-sided SEP showed a trend toward recovery.
Marked burst EMG activity was observed in the left gastrocnemius and tibialis anterior muscles during removal of the herniated nucleus pulposus.

At the end of the procedure, the operative field was thoroughly irrigated. The surgical team confirmed that there was no active bleeding and that the nerve root had been completely released.
MEP and SEP were then monitored again. MEP showed no substantial change compared with the waveform recorded during endoscopic exploration. The left-sided SEP remained unchanged, while the right-sided SEP had partially recovered compared with its intraoperative value.

Intraoperative monitoring during endoscopic lumbar discectomy places important demands on anesthetic management. Anesthetic agents can affect neurophysiological signals, so anesthetic management should be planned in advance with the anesthesiologist to ensure accurate monitoring.
Neuromuscular blocking agents should be avoided whenever possible because they may suppress EMG and MEP waveforms. SEP is susceptible to suppression by inhalational anesthetics. Total intravenous anesthesia is therefore preferred, with stable light-to-moderate anesthesia maintained throughout the procedure.
Because the interval from induction to the beginning of surgery is short in this type of procedure, the induction dose of neuromuscular blocker should be limited to the amount required for intubation. This allows an assessable MEP waveform to be obtained as early as possible before or soon after the start of surgery.
Continuous monitoring of EMG, SEP, and MEP across different time points and surgical maneuvers can reflect the effect of the procedure on neural function. When a maneuver such as traction or electrocautery directly stimulates a nerve root, abnormal EMG activity, including burst potentials, may appear. The surgeon should immediately stop the relevant maneuver and adjust the surgical trajectory to avoid potentially irreversible neural injury.
In this case, the left-sided SEP did not decrease. The right-sided SEP decreased during endoscopic exploration, but the decrease did not exceed 20%. The right-sided response subsequently showed a recovery trend through closure, and the final right-sided SEP was only 8.7% below the preoperative baseline.
The MEP waveform did not decrease, and latency did not show prolongation. These findings were interpreted as indicating that the patient’s sensory and motor conduction pathways were unlikely to have sustained significant injury. The procedure achieved the intended objectives of precise nucleus-pulposus removal, nerve-root release, minimally invasive decompression, and reduction of pain and weakness, with the aim of improving quality of life.
Blood pressure, heart rate, and body temperature should be maintained within stable ranges. Marked blood-pressure fluctuations may compromise perfusion of the brainstem or neural tissue and affect neurophysiological signals. Intraoperative temperature should be strictly controlled because hypothermia may slow nerve-conduction velocity and interfere with monitoring indices.
Electromagnetic interference from electrocautery should also be minimized to ensure that the monitoring equipment functions properly. Artifact control is an essential part of monitoring because it helps the team exclude changes unrelated to the surgical maneuver.
Endoscopic lumbar surgery is performed in a confined working space, with the nerve root often immediately adjacent to the operative field. Because the patient cannot provide spontaneous pain feedback under general anesthesia, continuous multimodal monitoring with SEP, MEP, and spontaneous EMG can provide real-time warning of possible nerve-root or spinal cord injury and may help reduce the risk of postoperative numbness, weakness, or paralysis.
In this approach, EMG provides immediate warning of direct nerve-root stimulation during close endoscopic manipulation. The combination of SEP and MEP provides complementary monitoring of sensory and motor function.
This case illustrates the value of multimodal IONM during endoscopic lumbar discectomy: patient-specific baselines, anesthesia compatible with monitoring, continuous waveform assessment, prompt response to burst EMG, and comparison of final signals with the preoperative baseline. These elements support minimally invasive decompression while helping protect neurological function.